SWAN Analytical A-13.411.100 Conductivity and Resistivity Meter

- +
Custom Settings Company Homepage More Products

The quotes are provided by Several Partner Companies

  • Prices: Request for a Quote for Further Details
  • Model : A-13.411.100
  • Category: Electrical Conductivity & Electrical Resistivity Meter
  • Brand : SWAN Analytical USA, INC.
  • Delivery Time: Request for a Quote for Further Details
  • Manufacturer: SWAN Analytical

Products

Features

  • Measurement and control transmitter in rugged aluminum housing (ip66).
  • For connecting four-electrode conductive sensors with integrated pt1000 temperature sensor,such as swansensor shurecon p or swansensor shurecon s.
  • The measurement range is from 0.05μS/cm to 100 mS/cm.
  • Used to measure specific conductivity,concentration (nacl,naoh and acid),salinity (in nacl) and total dissolved solids (tds in %).
  • Large backlit LCD screen,which can read measured values,sample temperature,sample flow rate and working status.
  • Simple user menu in English,German,French and Spanish. Simple programming of all parameters via keypad.
  • Galvanically isolated sensor connections.
  • According to en / din,natural water temperature compensation can be selected as a coefficient or a non-linear function.
  • Input and output overvoltage protection.
  • Two current signal outputs for measurement signals (0/4 - 20 ma).
  • Potential-free alarm contacts serve as programmable alarm values and summary alarm indications of instrument faults.
  • Two potential-free contacts can be programmed as limit switches or pid controls.
  • Input potential-free contacts to freeze measured values or interrupt control in automatic devices (holding function or remote shutdown).
  • The measuring and control transmitter has a robust aluminum housing (IP 66).
  • Measurement range from 0.05 μS/cm to 100 mS/cm.
  • Simple user menu at English,German,French and Spanish. Simple programming of all parameters by keyboard.

Electron emitter/controller for measuring CIP solution specific conductivity,concentration,salinity and TDS,characteristics

Product Specifications

Parameter Value Parameter Value
precision 0.5000 ±% Full Scale Medium temperature -22 to 482 F (-30 to 250 C)
The number control relay 2 The electrical output Analog Current
Display type Digital Display User control Digital Front Panel
Working temperature 14 to 122 F (-10 to 50 C) Appearance of the form Panel
Conductivity range 0.0500 to 100 µS/cm Features and functions Automatic Temperature Compensation
Maximum input channel 1 (# channels) Product category Conductivity and Resistivity Meters
The water quality parameters Measures Conductivity / Dissolved Solids (optional feature)