Messkonzept FTC320 Ex online explosion-proof thermal conductivity analyzer gas instrument

- +
Custom Settings Company Homepage More Products

The quotes are provided by Several Partner Companies

  • Prices: Request for a Quote for Further Details
  • Model : FTC320 Ex online explosion-proof thermal conductivity analyzer
  • Category: Gas Equipment
  • Brand : Xincheng Technology
  • Delivery Time: Request for a Quote for Further Details
  • Manufacturer: Xincheng Technology

Products

Features:

l Corrosion-resistant protection: The use of anti-corrosion flow paths and polymer coatings can measure corrosive gases;

l Optional humidity measurement module to avoid cross-interference caused by humidity fluctuations;

l Condensation water and dust protection: using special micron filters;

l Equipped with cross-interference gas compensation function;

l Can be used for combustible gas measurement;

l Multi-gas measurement mode;

l Configurable integrated flow and pressure measurement modules;

l Explosion-proof level: EX-ZONE2 EX II3G NR IIC T4 GC


Measuring components and range:


Basic measuring rangeMinimum measuring rangeGas componentsmulti-gas mode
H2(hydrogen)0-100%0-0.5%Nitrogen or airsupport
O2 (oxygen)0-100%0-15%Nitrogensupport
0-100%0-0.8%Nitrogen or airsupport
CO2(carbon dioxide)0-100%0-3%Nitrogen or airsupport
N2 (nitrogen)0-100%0-3%Argonsupport
O2 (oxygen)0-100%0-2%Argonsupport
H2(hydrogen)0-100%0-0.4%Argonsupport
0-100%0-0.5%Argonsupport
CO2(carbon dioxide)0-60%0-10%Argonsupport
Ar (argon gas)40%-100%carbon dioxidesupport
CH4(methane)0-100%0-2%Nitrogen or airsupport
CH4(methane)0-100%0-1.5%Argonsupport
Ar (argon gas)0-100%0-3%oxygensupport
N2 (nitrogen)0-100%0-2%hydrogensupport
O2 (oxygen)0-100%0-3%carbon dioxidesupport
H2(hydrogen)20%-100%20%-40%Helium
H2(hydrogen)0-100%0-0.5%Methane
H2(hydrogen)0-100%0-0.5%carbon dioxide
SF6 (sulfur hexafluoride)0-100%0-2%Nitrogen or air
NO2 (nitrogen dioxide)0-100%0-5%Nitrogen or air
H2(hydrogen)0-100%0-0.8%oxygen
Ar (argon gas)0-100%0-3%xenon
Ne(neon)0-100%0-1.5%Argon
Kr(Krypton gas)0-100%0-2%Argon
Fire extinguishing agent (R125)0-100%0-5%Nitrogen or air
D2(deuterium)0-100%0-0.7%Nitrogen or air
D2(deuterium)0-100%0-5%Helium


Specifications:


Measurement principleThermal conductivity
Sampling methodSampling
Preheat time20min; small range 60min
Sample flow10 l/h - 150 l/h, 60 l/h - 80 l/h (recommended)
Sampling tube6mm
T90(1L/min)<1sec
digital communicationRS232
Analog communication0/4-20 mA;0-10V
measurement accuracy1%FS
Noise interference<1% of minimum range
zero drift<2% of minimum range
Repeatability<1% of minimum range
nonlinear<1% FS
Measurement error (ambient temperature changes by 10 degrees)<1% of minimum range
Flow rate (1-1.5L/min, per 0.16L/min)<1% of minimum range
Pressure effect (Pabs >800 hPa) per 10 hPa)<2% of minimum range
Dimensions (WxHxD mm)145x80x85
Protection levelIP65
Pressure resistantMaximum 20 bar (absolute pressure)
Power supply18V to 35V DC
Working temperature‘-10°C to 50°C
weight1800g


Product Specifications

Parameter Value Parameter Value
Response Time 10seconds The channel number 1 channel
Characteristics of the Dangerous environment; The electrical output Analog current;
According to digital Working temperature 0-50C
Working humidity 0-80% Installation type fixed
Measuring type % lower explosive limit; Type of gas Combustible gas;
The instrument type Analyzer application Process control gas monitoring
Monitoring gas quantity A variety of gas;

Download

MK Gas Analysis Product Application Areas 2022