Bodong Pfeiffer Residual Gas Analyzer PrismaPro Residual Gas Analyzer
- Prices: Request for a Quote for Further Details
- Model : Residual gas analyzer PrismaPro
- Category: The Residual Gas Analyzer
- Brand : Badong Enterprise
- Delivery Time: Request for a Quote for Further Details
- Distributor: Badong Enterprise
Products
Germany Pfeiffer Residual gas analyzer (Quadrupole mass spectrometer) integrates high sensitivity,high stability and intelligent operation. It is equipped with a new generation of operating software PV MassSpec,which quantitatively and qualitatively evaluate the residual gas components in the vacuum system. The lowest detectable partial pressure is 3 X10-15hPa.
Pfeiffer Residual Gas Analyzer PrismaPro Features:
Modular design ensures compatibility,JSON programming interface and multiple interfaces facilitate system integration
Analytical instruments and electronic units are flexibly interchangeable
Dual filament extends working time
Minimum detectable partial pressure 3 X10-15hPa
Can be directly connected to a vacuum gauge to facilitate pressure measurement
Pfeiffer residual gas analyzer PrismaPro technical parameters:
model | QMG 250 F1 | QMG 250 F2 | QMG 250 F3 | QMG 250 M1 | QMG 250 M2 | QMG 250 M3 | |
Detector | Faraday (F) | Electron Multiplier / Faraday C-SEM / Faraday (M) | |||||
mass number amu | 1–100 | 1–200 | 1–300 | 1–100 | 1–200 | 1–300 | |
Quadrupole diameter/length | 6 / 125 mm | ||||||
Minimum detection limit F hPa *1,2 | 4X10-13 | 5X10-13 | 7X10-13 | - | - | - | |
Minimum detection limit M hPa *1,2 | - | - | - | 3X10-15 | 4X10-15 | 5X10-15 | |
Sensitivity to Ar F A/hPa*3 | 5X10-4 | 4X10-4 | 3X10-4 | 5X10-4 | 4X10-4 | 3X10-4 | |
Maximum working pressure F hPa | 5X10-4 | ||||||
Maximum working pressure M hPa | - | - | - | 5X10-5 | 5X10-5 | 5X10-5 | |
Effect on adjacent masses*1 | < 10 ppm | < 20 ppm | < 50 ppm | < 10 ppm | < 20 ppm | < 50 ppm | |
Operating temperature analysis | 200 °C (max. 150 °C when operating with SEM) | ||||||
Operating Temperature Electronics | 5 – 50 °C | ||||||
Baking temperature analysis | 300 °C | ||||||
connecting flange | DN 40 CF-F | ||||||
compress time | 1 ms – 16 s/amu | ||||||
weight | 2.5 kg | 3.2 kg |
*1open ion source,
*2Residence time 4 seconds
*3Compared with C-SEM,has higher sensitivity
Pfeiffer residual gas analyzer PrismaPro standard communication interface
Pfeiffer Residual Gas Analyzer PrismaPro Scalable communication interface:
Pfeiffer residual gas analyzer structure video