Shuide Technology Shallow Water Sensor
- Prices: Request for a Quote for Further Details
- Model : Shallow water type
- Category: Water Quality Testing Instrument
- Brand : Shuide Technology
- Delivery Time: Request for a Quote for Further Details
- Distributor: Shuide Technology
Products
Technology In-situ Nutrient Analyzer
Category: On-site nutrient salt monitor
Model: Shallow water type
Keywords: in-situ nutrient analyzer,nutrient analyzer
Supplier: Technology Co.,Ltd.
Product introduction:
The in-situ nutrient analyzer is a chemical sensor suitable for shallow waters. It can sequentially measure dissolved ammonia nitrogen (NH3-N),nitrite (NO2-N),nitrate (NO3-N),and phosphoric acid in the water body. Five nutrients: salt (PO4-P) and silicate (SiO3-Si). It is small in size,easy to carry and deploy,and easy to install on monitoring platforms such as buoys,shore stations,and survey ships. It is suitable for water bodies such as oceans,estuaries,rivers,lakes,etc.,and is used for eutrophication research,phytoplankton growth research,and environmental monitoring. Provide high-precision,continuous and stable data.
In-situ nutrient analyzer
In-situ nutrient analyzer
The in-situ nutrient analyzer is a chemical sensor suitable for shallow waters. It can sequentially measure dissolved ammonia nitrogen (NH3-N),nitrite (NO2-N),nitrate (NO3-N),and phosphoric acid in the water body. Five nutrients: salt (PO4-P) and silicate (SiO3-Si). It is small in size,easy to carry and deploy,and easy to install on monitoring platforms such as buoys,shore stations,and survey ships. It is suitable for water bodies such as oceans,estuaries,rivers,lakes,etc.,and is used for eutrophication research,phytoplankton growth research,and environmental monitoring. Provide high-precision,continuous and stable data.
characteristic:
With pre-filter,it can adapt to water with high turbidity;
Modular design makes it easy to replace accessories and maintain;
Equipped with waste liquid recovery device to avoid environmental pollution;
Comes with data storage backup function;
Can be automatically corrected.
Technical specifications:
nitrite | Nitrate | Phosphate | silicate | Ammonia nitrogen | |
Analytical method | Diazo-Azo method | UV/Cadmium column reduction Diazo-Azo method | Phosphomolybdenum blue method | Silicon molybdenum blue method | Fluorescence method |
Detection wavelength | 543 nm | 543 nm | 880 nm | 810 nm | 365 nm /420~470nm |
Measuring range(μg/L) customizable | 0-200 | 0-2000 | 0-500 | 0-2000 | 0-500 |
Precision | 5% | ||||
Accuracy | ±10% | ||||
Maintenance cycle | greater than 2months | ||||
working environment | Water depth: 0~10m; Water temperature:5~40℃ | ||||
volume | Host: 160mm(diameter)×580mm(high) | ||||
Material | Host shell: hard PVC; Fixed bracket: stainless steel | ||||
Voltage | 12V DC | ||||
Power consumption | average less than 10w | ||||
Communication Interface | RS232/485 |
More keywords:In-situ nutrient salt analyzer,nutrient salt analyzer,domestically producedNutrient analyzer